Calibration of Digital Thermometer using Sensor and Indicator Method
Fitria Hidayanti1, Ukhti Fathimah Ramadhani2
1Fitria Hidayanti*, Engineering Physics Department, Universitas Nasional, Jakarta, Indonesia.
2Ukhti Fathimah Ramadhaniedi, Engineering Physics Department, Universitas Nasional, Jakarta, Indonesia.
Manuscript received on April 25, 2020. | Revised Manuscript received on May 04, 2020. | Manuscript published on May 30, 2020. | PP: 11-14 | Volume-4, Issue-9, May 2020. | Retrieval Number: I0838054920/2020©BEIESP | DOI: 10.35940/ijmh.I0838.054920
Open Access | Ethics and Policies | Cite
© The Authors. Published By: Blue Eyes Intelligence Engineering and Sciences Publication (BEIESP). This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/)
Abstract: A Calibration has been done for digital thermometer using a sensor and indicator method. The unit under test manufactured by Fluke with 54-II type and serial number 13160058, meanwhile, the standards that used are PT100 sensor from Omega and an indicator from Additel. The calibration result from 12 setpoints shows the average correction value as 0.61 ºC with uncertainty as 0.32 ºC and K factor as 2.0 at 95% confidence level.
Keywords: Calibration, correction, digital thermometer, standard, uncertanity.